Calibration Procedure for Pharmaceutical Industries
1.
Scope :- This procedure applies to the Process/
Manufacturing and laboratory equipment used by the Pharmaceutical Industries.
2.
Purpose :- The
Purpose of procedure specifies the
schedule and requirements for calibration, performance verification, and
maintenance of Laboratory, Process/ Manufacturing testing instruments and equipment.
3.
Roles and Responsibility :-
A.
User Department :- User Department Ensure the Equipment/
Instrument Availability of the System as Desired Instrument List.
B.
QA Engineering Department: - is
Responsible for Preparing, Implementing and Maintaining the Procedure. And
Review the Instrument/ Equipment Planner and Check the Calibration Row Data and
Calibration Certificate as per Desired Point.
C.
Calibration Owner: - is
Responsible for Following the Procedure as
per Approve Procedure.
D.
QA Department: - QA Department is Responsible for Assigned
Calibration Owner and User Department responsibility as per the Defined
Procedure.
4.
Terms and Definition
4.1
Calibration :- Calibration is Procedure for Adjustment or
standardization of the accuracy of a measuring instrument, usually by
comparison with a certified reference or standard.
4.2
Traceability:- The
linking of measurement standards and/or measuring instruments to relevant
national or international standards through an unbroken chain of comparisons.
4.3
Like to Like :- The instrument which has same specification
like Least Count, Accuracy, Range, and Process connection.
4.4
Calibration Range :- Measurement range of the
instrument for which calibration has been / is to be carried out.
4.5
Least Count :- Measurable Small Unit is Least Count.
4.6
Out of Calibration:- Instrument
is Working Condition and Found Instrument Out Side the Accuracy ,Beyond the
Limit.
4.7
Criticality Assessment :- He process of evaluating the impact of
operating, controlling, alarming and failure conditions of a system on the
quality of a product Basically Instrument Divided into Three Category a.
Critical b. Non Critical c. For Information
4.8
Critical Instrument: - Impact on Direct Production Process.
4.9
Non
Critical Instrument :- No
Impact on Production Process.
4.10
For Information :- No Directly and Indirectly Impact on
Production Process, Only on Only Indicative Purpose.
5.
Procedure:-
Adding Procedure for New Instrument/Equipment
5.1
Rise the Change Control Procedure: - Defined the Why /Where We are Install a
Equipment and General Aspect full fill as per Approved Change Control
Procedure.
5.2 Criticality
Assessment: - Defined the
Instrument Criticality Assessment as Desired Process Parameter. After
Assessment Defined the Instrument category.
5.3
Assign the Equipment and Instrument Identification
Number. After Installation of
Equipment and Instrument as Per Criticality Assessment Assigning the Unique
Identification no. for Each and Every Instrument. Numbering System Assign as
per Equipment Numbering Sop.
5.4 Defined
Calibration Frequency: - Defined the
Calibration Frequency as Per Criticality Assessment. General Terms of Define
the Frequency
Sr. No. |
Instrument Category |
Frequency |
Tolerance |
1. |
Critical |
6 Month |
+/- 30 Days |
2 |
Non Critical |
12 Month |
+/- 30 Days |
3 |
For Information |
No Need As
Requirement |
NA |
4 |
Master
Instrument |
12 Month |
+/- 60 Days |
5.6 Prepare
Calibration Master Plan :- After identification of the Instrument and
Categorization and Frequency That
Instrument Shall be Add in Calibration
Master Plan. CML Should be Revised in Every Once in Year.
5.7 Prepare
Calibration Schedule :- The
schedule which defines the frequency after which the instrument should be
calibrated.
6.
DETAILS OF PROCEDURE
6.1 General Equipment
Requirements
The
QA Manager or designee shall maintain an equipment inventory that shall include
the following information:
A.
Sr. No
B.
Instrument ID
C.
Instrument Location
D.
Accuracy
E.
Acceptance Criteria
F.
Category
G.
Remark
H. History
I.
Remark
6.1.1
If any Instrument
Added and Delete and Modified, Prepare the Addendum of Instrument Manufacturing
List With Respect to Change Control
6.1.2
Instrument List
Shall be Revised in Once in a year.
6.1.3
All equipment in
the equipment inventory system shall be identified uniquely.
6.1.4
All equipment
shall be maintained in good operating order and according to the manufacturer
and/or Section maintenance requirements.
6.1.5
All critical
equipment shall be calibrated or verified before use.
6.2 Master Standards
6.2.1
Whenever
possible, Master standards traceable to SI units (International System of
Units) shall be used. In situations where SI units cannot be used, certified
reference material provided by a competent supplier shall be used if available.
6.2.2
Master
standards shall be calibrated by an accredited organization or vendor that can
provide proof of traceability. These typically would include, but not be
limited to, ISO 17025-certified companies.
6.2.3
Reference
standards shall only be handled by employees authorized by the QA
Manager/Supervisor and shall be stored to prevent contamination and/or
deterioration. Reference standards shall be calibrated before and after any
adjustment. All reference standards, certified reference materials, or
reference materials used for calibration shall be uniquely identified. A
certificate of traceability, if applicable, shall be retained to ensure
traceability.
6.2.4
QA
Managers/Supervisors shall include procedures for calibration and/or
performance verification of new equipment in Section technical procedures.
6.2.5
calibration
procedures shall be appropriate for the intended use of the equipment and shall
provide criteria for determining if calibration is satisfactory.
6.2.6
Calibration records shall
be maintained and associated with the unique identifier of each piece of
equipment. These records shall include:
·
the identity of the item of equipment.
·
Name of manufacturer.
·
Serial number or unique identifier.
·
Date of calibration.
·
Current location.
·
Manufacturer’s instructions for Useable Instrument
·
The reference standard certified reference material or reference
material used for calibration.
·
Copies of all reports, results of calibration, and/or
certificates of calibration.
·
A maintenance plan and due date for the next calibration.
·
The identity of the individual performing calibration.
6.2.7
When
external calibrations are performed, service providers that demonstrate
competence, measurement capability, and traceability shall be used. Calibration
certificates from these providers shall contain the measurement results,
including the measurement uncertainty and/or a statement of compliance with an
identified metrological specification. When possible, providers accredited to
ISO 17025 shall be used. Copies of the provider’s accreditation documentation
shall be maintained by the Quality Engineering (QE).
6.2.8
If
calibrations are performed by an outside vendor, the Section document control
custodian shall maintain the original calibration records provided by the
vendor and a copy of the relevant records shall be readily available.
6.2.9
Laboratory
equipment requiring calibration shall be labelled or coded to indicate the
calibration status, including the date when last calibrated and the due date
for recalibration (or expiration criteria for when recalibration is due).
6.3 Out of Service Instrument :-
6.3.1
Equipment that
has been subjected to overloading or mishandling, gives suspect results or has
been shown to be defective or outside specified limits, shall be taken out of
service. It shall be isolated and/or clearly labelled (Out of Service – Do Not
Use) to prevent use until repaired and shown by calibration or test to perform
correctly.
6.3.2
Prior to
returning a piece of equipment to use (out of service for any reason – e.g.,
maintenance, malfunction, leaving the direct control of the Laboratory), the
correct operation shall be demonstrated by calibration or performance
verification. Laboratory personnel shall examine the effect(s), if any, of a
malfunction on analysis results and implement the Procedure for Corrective
Action as required.
6.4
out of
schedule:-
6.4.1
If calibration
is not performed within the scheduled due date, an Incident notification shall
be raised & investigation shall be carried out and details of
investigation shall be documented.
6.5
unavailable
for calibration :-
6.5.1
If an
equipment/instrument is under validation/Major maintenance/Breakdown etc. then
Calibration shall not be performed & calibration order shall be locked with
reference document / notification number.
6.6
Adding/Deleting/Removing/Modification :-
6.6.1
If any Case
Instrument Shall be Removed from
Equipment, CML to be Update as per Desired Format and Also Mention the Remark
Reference Change Control or Document No. And Also Update the Calibration
Schedule.
6.6.2
If
any Case Instrument Shall be Add
from Equipment, CML to be Update as per Desired Format and Also Mention the
Remark Reference Change Control or Document No. And Also Update the Calibration
Schedule.
6.6.3
If
any Case Instrument Shall be Modify
(Range, Location, Least Count ..Etc) from Equipment, CML to be Update as
per Desired Format and Also Mention the Remark Reference Change Control or
Document No. And Also Update the Calibration Schedule.
Annexure-I Instrument
Criticaility Assesment |
||||||
Equipment Name |
|
|
||||
Equipment ID |
|
|
||||
Assesment Point |
Equipment ID |
Equipment ID |
||||
|
|
|||||
Impact |
Non Impact |
|
||||
Yes |
No |
Yes |
No |
|||
Does the system automate or control
any of: Manufacturing, Sterilization, Formulation, Labeling, Inventory, or
Critical Environment Controls? |
|
|
|
|
||
Is the system the original source
of data for the automation or control of any of: Manufacturing,
Sterilization, Formulation, Labeling, Inventory, or Critical Environment
Controls? |
|
|
|
|
||
Is this system used to generate,
manage and analyze data concerning Product Quality, Safety, Efficacy,
Strength Stability or identify? |
|
|
|
|
||
Is this system used for raw and
in-process material, clinical data analysis, automated inspection equipment
and laboratory data system? |
|
|
|
|
||
Is this system used for cGxP
Support Functions like Calibration, Maintenance Scheduling, Quality Trending? |
|
|
|
|
||
Is this system used for market
complaints or adverse event reporting? / electronic document
submission/reporting to regulatory agencies |
|
|
|
|
||
Is this system used to handle CAPA,
Change Control, Incident, Audit, testing, lab event? |
|
|
|
|
||
Is this system used to maintain
records of training of personnel? |
|
|
|
|
||
Is the system used to maintain
copies of protocol pertaining to non-clinical study? |
|
|
|
|
||
If One
Point are yes Instrument is Critical Otherwise Instrument Will be Non Critical.
4 Comments
Good job ashish for sharing your technical knowledge
ReplyDeleteThanks for Appreciation. Stay tuned with us.
DeleteGreat article by the great author, it is very massive and informative but still preaches the way to sounds like that it has some beautiful thoughts described so I really appreciate this article. Best Valve testing in UAE Services Provider
ReplyDeleteReally Helpful, I would like say thanks for the whole team.
ReplyDelete